CoCo-80 User Manual
110
Figure 123. Overlap rate for transient capture.
After the Parameter Settings are specified the CoCo begins to wait for a trigger
event. A window displays the time elapsed before a trigger event is detected.
Figure 124. Waiting for trigger message.
When a trigger event is detected and Manual-Arm Trigger is selected, then a small
window will show the data frame and give you the option to accept or reject the
data. Accept will include the frame into the average and then ask you to proceed
to the next trigger by pressing the Next button. Reject will discard the frame, not
include it in the average and return to the waiting for trigger mode. If Auto-Arm
Trigger mode is selected then the system will automatically return to the wait
mode after each trigger event with no user intervention.