CoCo-80 User Manual
37
Figure 39: The sensor parameter menu
Measurement Quantity defines the quantity such as acceleration,
velocity, displacement, force, voltage, etc.
Sensor Engineering Units defines the engineering units such as
m/s
2
, cm/s
2
, g, etc. for the input channel.
Sensor Sensitivity defines the sensitivity in
millivolts/engineering unit defined in the unit menu. This
selection opens a numeric keypad to enter the sensitivity value. Press
the OK soft button to accept the value.
dB Reference defines the dB reference value for certain
measurement quantities.
Select Integration/Differentiation gives the options of No
Integration/Differentiation, Integration/Differentiation, or Double
Integration/Differentiation when the measurement quantity is
acceleration, velocity, or displacement.
When the Physical Quantity is set as Acceleration a built-in integration or double-
integration module may be used to generate readings in velocity or displacement.
When the Physical Quantity is selected as Velocity, integration to displacement or
differentiate to acceleration modules can be enabled. Note that the algorithms for
integration are implemented in the digital domain and include a high-pass filter
and DC removal routines.