I-Class Mark II Maintenance Manual
Ad
j
ustments and Maintenance
Standard Calibration (continued)
Step Action Displayed Message Comment
F
Proceed according to the media
type:
All media except Continuous –
Position label material (and liner, if
any) over the sensor then press
the ESC button.
Continuous – Install media.
Position the Media Sensor under
the stock and press ESC.
SCAN PAPER
PRESS ESC BUTTON
yyy
This sets the paper value, where
“yyy” represents the current
sensor reading.
(1) If using preprinted media,
ensure that the area placed over
the sensor is free of text,
graphics, or borders.
(2) See the Operator’s Manual
for sensor adjustment
instructions.
G Observe the calibration result. CALIBRATION COMPLETE
Calibration was successful.
If WARNING LOW BACKING is
displayed, calibration was
successful.
H
Press ESC then EXIT to return to
READY.
When calibrating gap or reflective
media, press and hold the FEED
button until at least one label is
output.
CALIBRATION COMPLETE
Followed by...
READY
The printer is now ready for use.