calibration
result is not
updated
64-1 Note
ISE the
concentration
of internal
solution
abnormity
1the
concentratio
n of
internal
solutionC
IS))is
within
following
range :Na
CIS)<
120.0mmol/
l or
190.0mmol/
CIS
2current
ISE
calibration
result is not
updated
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
(3) Please re-execute ISE calibration.
64-2 Note
ISE the
concentration
of internal
solution
abnormity
1the
concentratio
n of
internal
solutionC
IS))is
within
following
range .KC
IS)<
3.0mmol/l
or 8.0mmol/
CIS
2current
IS
calibration
result is not
updated
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
(3) Please re-execute ISE calibration.
64-3 Note
ISE the
concentration
of internal
solution
1the
concentratio
n of
internal
solution
C
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device