EasyManua.ls Logo

Dirui CS-T240 - Page 108

Dirui CS-T240
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
108
ISE
calibration
result is not
updated
63-2 Note
ISE slope
value
(SLOPE)
abnormal
1The
slope value
of
calibration
result is
within
following
range or the
impact of
electrode is
low
(cross-conta
mination
rate (A) is
as
following:K
:(1
SLOPE
32mv
2current
ISE
calibration
result is not
updated
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
(3) Please re-execute ISE calibration.
63-3 Note
ISE SLOPE
abnormal
1The
slope value
of
calibration
result is
within
following
range or the
impact of
electrode is
low
(cross-conta
mination
rate (A) is
as
following:C
l:(1
SLOPE
-25.0mv
2current
ISE
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
(3) Please re-execute ISE calibration.

Table of Contents