ISE
calibration
result is not
updated
63-2 Note
ISE slope
value
(SLOPE)
abnormal
(1)The
slope value
of
calibration
result is
within
following
range or the
impact of
electrode is
low
(cross-conta
mination
rate (A) is
as
following:K
:(1)
SLOPE<
32mv
(2)current
ISE
calibration
result is not
updated
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
(3) Please re-execute ISE calibration.
63-3 Note
ISE SLOPE
abnormal
(1)The
slope value
of
calibration
result is
within
following
range or the
impact of
electrode is
low
(cross-conta
mination
rate (A) is
as
following:C
l:(1)
SLOPE>
-25.0mv
(2)current
ISE
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
(3) Please re-execute ISE calibration.