Chapter 3: Operation
Operation Manual: Evolution Series
Command Response Function
[AA CONT:MOTO:STAT?] [ON] or [OFF] Report the status of motor.
[AA CONT:MOTO:SET?] [NNN] Report the setting of RPM (NNN).
[AA CONT:HEAT:STAT ON] [ON] Turn the heaters on.
[AA CONT:HEAT:STAT OFF] [OFF] Turn the heaters off.
[AA CONT:TEST:ABORT] [!] Abort the current test.
[AA CONT:TEST:CLR] [!] Reset the delta.
[AA CONT:TEST:START] [!] Start the current activated method.
[AA MEAS:ALL] [a1,a2, ….a11]
Report rpm, max rpm, min rpm, temp at V1,
V2 … V8.
[AA MEAS:RPM] [NNNN] Report RPM (NNN.N).
[AA MEAS:RPM:MAX] [NNNN] Report max RPM (NNN.N).
[AA MEAS:RPM:MIN] [NNNN] Report min RPM (NNN.N).
[AA MEAS:TEMP:AVG] [NNNN]
Report the averaged temperature of all
enabled vessels
[AA MEAS:TEMP:CUR X] [NNNN] Report the current temperature of vessel X.
[AA MEAS:TEMP:MAX X] [NNNN] Report the max temperature of vessel X.
[AA MEAS:TEMP:MIN X] [NNNN] Report the min temperature of vessel X.
[AA MEAS:TEMP:MAX] [a1,a2, ….a7]
Report the max temperatures of vessels 1, 2 …
7.
[AA MEAS:TEMP:MIN] [a1,a2, ….a7]
Report the min temperatures of vessels 1, 2 …
7.
[AA SYST:DATE YYYY,MM,DD] [YY,MM,DD] Set the system date.
[AA SYST:TIME HH,MM,SS] [HH,MM,SS] Set the system time.
[AA SYST:DATE?] [YY,MM,DD] Report the current date.
[AA SYST:TIME?] [HH,MM,SS] Report the current system time.
[AA SYST:CAL?]
[iiii,ssss,iiii,ssss,
…,iiii,ssss]
Report cal settings: intercept iiii and slop ssss
of each vessel.
[AA SYST:ETIME?] [HH,MM] Report the elapsed time as hours and minutes.
[AA SYST:ADDR?] [AA] Report the address of the unit.
[AA SYST:SLEEP] [!] Puts the system in sleep mode.
[AA SYST:ADDR AA,PP,TT] [!] Set the addresses for the unit, printer and TCS
[AA SYST:PRNT:REQ BB] [PRNT DD]
The unit BB requests a printer service from unit
AA. Valid DD responses:
• 0: the request is allowed
• 1: the printer of the unit AA is off or
offline
• 2: the printer of the unit AA is out of
paper
• 3: the printer of the unit AA has an
error
• 4: the printer of the unit AA is busy.
Please wait
• 5: the unit AA doesn't have a local
printer