Chapter 3: Operation
Operation Manual: Evolution Series
Command Response Function
[BB NEXT] [ssssss …..sssss]
Tell the unit BB to send a string to print.
Where sssss…sssss is a string (must be less than
81 characaters) to be printed.
[AA *IDN?]
[6100/6300,SSSSSSS,PVRev
NNN]
Report the bath type, serial number (SSSSSSSS)
and Firmware Rev (NNN).
[AA HCOP:IMM] [!] Print an instant report.
[AA PKGT:METH:HEAD
NNNNNNNNNNNNNNNN
HHHH MM
HHHH MM
DDD
S
TTT
RRR
iii
LLL
E
P
[!]
Set the header of a new method and saves the
header. The following variables can be set:
• Name: NNNNNNNNNNNNNNNN
• Test Length: HHHH (hours) MM (min.)
• IRPM Length: HH (hours) MM (min.)
• Delayed Drop: DDD seconds
• Report Switch: 0 (Off) or 1 (On)
• Temperature: TT.T°C
• RPM: RRR
• IRPM: iii
• Volume: LLL mls
• Enable Vessels: E: Bit X: Vessel X (1:
enable, 0: Disable)
• Stirrer Type: P for PADDLE & B for
BASKET
[AA PKGT:METH:SAMP l,
HHHH MM
HHHH MM
HHHH MM
HHHH MM
HHHH MM
HHHH MM
HHHH MM
HHHH MM]
[!]
Set the list of sample time points (from i*8)
and saves the list. The max number of time
points of each command is 8. Total number of
time points of each method is 16
[AA PKGT:METH:HEAD?]
[NNNNNNNNNNNNNNNN
HHHH MM]
Report the method header.
[AA PKGT:METH:SAMP? l] [HHHH MM] Report the time points from l*8.