Operation
Manual
Evolution
Dissolution Test
System
6100
6300

Table of Contents

Questions and Answers

  • Z
    Zachary LopezSep 10, 2025
    Why does the main circuit keep tripping on my Distek Evolution 6100 Test Equipment?
    • K
      Kenneth Ray MDSep 12, 2025
      If the main circuit of your Distek Test Equipment keeps tripping, it is likely due to a short circuit within the unit. Inspect the device for any shorts, starting with the power switch, transformer, wiring, and system circuit boards.
  • P
    patrick51Sep 12, 2025
    What causes no +6V and no -5V on Distek Evolution 6100?
    • J
      John DiazSep 12, 2025
      If your Distek Test Equipment shows no +6V and no -5V, measure the output voltage coming from the power supply at +12V and -12V. Check that the wiring is securely soldered into the power supply terminals and the connector that supplies the interface board (5150) for any loose connections. If the problem continues, replace the defective interface board (5150).
  • B
    Becky PierceSep 14, 2025
    What to check if there is no low voltage AC from the transformer in Distek Evolution 6100 Test Equipment?
    • D
      Danny RussellSep 14, 2025
      If your Distek Test Equipment has no low voltage AC (secondary) from the transformer, measure if there is input voltage coming into the transformer primaries (115v or 2202v). Also, check that the transformer wirings are securely fastened into the I/O connector.
  • L
    Lydia MannSep 16, 2025
    What to do if there is no high voltage (HV) on Distek Evolution 6100?
    • S
      Stephanie ConnerSep 16, 2025
      If there is no high voltage (HV) on your Distek Test Equipment, measure if there is output voltage coming out of the transformer secondary. If there is no AC voltage coming out of the secondary, replace the transformer. Check that the transformer wiring is securely soldered into the transformer terminals and the connector that supplies the motor/heater board for any connections. If the issue persists, replace the defective interface board (5150).

Summary

Introduction to Distek Evolution Dissolution System

System Specifications

Lists the technical specifications for the Evolution 6100 and 6300 models.

Operation Warnings and Notes

Highlights critical safety warnings and operational notes.

Installation Guide for Distek Evolution System

Leveling the Evolution System

Procedure for ensuring the instrument is level on the bench.

Paddle and Basket Installation

Detailed steps for installing paddles, baskets, and checking their alignment.

Operation of the Distek Evolution Dissolution System

Temperature Calibration

Procedures for calibrating the system's temperature sensors.

Method Management (Automatic Mode)

Guidance on selecting, sorting, and programming test methods.

Using the produKEY

Operations related to saving, viewing, and managing methods via produKEY.

Specific Calibration Procedures

Detailed steps for various temperature calibration methods.

Maintenance of the Distek Evolution Dissolution System

Troubleshooting Common Issues

Troubleshooting Overview and Safety

General precautions and initial steps before troubleshooting.

Error Messages

Lists potential error messages and their solutions.

Warning Messages

Explains system warnings and required user actions.

Display and Keypad Issues

Troubleshooting steps for LCD display and keypad malfunctions.

Power Supply Issues

Diagnosing problems related to the main and DC power supplies.

Heating and Speed Issues

Solutions for abnormal heating profiles and speed problems.

I;O Communication Issues

Troubleshooting printer, RS232, and RS485 communication problems.

Temperature and Calibration Issues

Resolving problems with temperature displays and calibration.

Mechanical System Issues

Addresses problems with leveling, shaft wobble, centering, and drive assembly.

Appendix B: Pre-Installation Electrical Power Considerations

Distek Evolution 6100 Specifications

General IconGeneral
ModelEvolution 6100
TypeDissolution Tester
Number of Stations6
Number of Vessels6
Data LoggingYes
ComplianceUSP, EP, JP
SoftwareDissolution Software
Vessel Volume1L or 2L
Stirrer Speed Range25 to 250 rpm
Basket Speed Range25 - 250 RPM
Paddle Speed Range25 - 250 RPM
Heater Power1000 W
DisplayLCD
Sampling CapabilityManual or automated
ConnectivityUSB, Ethernet
Power Requirements100-240V, 50/60Hz
Sampling PortsYes