Technical Data
102
Instructions for use D-Vapor/D-Vapor 3000
Filling time for one anesthetic agent bottle of des-
flurane (240 mL)
At 22 °C (71.6 °F) and with the vaporizer not
yet warmed up
1minute
At 22 °C (71.6 °F) and with a warmed-up va-
porizer, the filling time is increased.
<2 minutes
Flow resistance at a fresh-gas flow of 10 L/min
(STPD)
<100 hPa (102.2 cmH
2O)
Tilt angle
During operation 10°
During storage and transport No limit
Warm-up time at 22 °C (71.6 °F) approx. 5 to 6 minutes (depending on the input volt-
age and the fill level of the vaporizer)
Bridging time when mains power supply absent maximum of 5 minutes (at a concentration of
6 Vol% maximum, fresh-gas flow 4 L/min and fully
charged battery)
Noise emission during operation <45 dB (A)
Pneumatic interfaces
Filling systems Baxter SAFE-FIL filling system for desflurane (Su-
prane)
Piramal Fill filling system for desflurane (Torrane)
Connections
D-Vapor DW-2000, Auto Exclusion, Selectatec S-2000,
permanent connection, ISO conical adapter
D-Vapor 3000 Auto Exclusion
Alternating fresh-gas flow <30/min
Static pressure (above ambient pressure) at the in-
terface
max. 200 kPa (2 bar)
Use of latex D-Vapor and D-Vapor 3000 are not made with nat-
ural rubber latex.
Performance characteristics (Cont.)