Preparation
44
Instructions for use D-Vapor/D-Vapor 3000
Vaporizer with Baxter SAFE-FIL filling system for desflurane (Suprane)
Starting the filling procedure
1 Unscrew the cap from the anesthetic agent
bottle.
2 Ensure that the bottle, bottle connection, and O-
ring on the bottle connection are not damaged.
3 Press the unlocking button (A) and withdraw the
sealing plug (B) from the filling opening.
4 Insert the anesthetic agent bottle (C) into the
filling opening. The anesthetic agent bottle will
engage in the filling opening with an audible
clicking sound.
5 Press the anesthetic agent bottle down and
keep it pressed down.
Anesthetic agent will flow into the vaporizer. The
contents of the bottle will show bubbles.
6 Watch the filling level in the sight glass. When
the maximum mark is reached, the flow stops
automatically.
7 If the Vaporizer is filled when it is warmed up,
the contents of the tank will also bubble. Briefly
interrupt the filling to check the fill level. To do
this, stop pressing down the anesthetic agent
bottle. Then press the bottle down again and
continue with the filling.
8 Fill the vaporizer no higher than the maximum
mark .
If the vaporizer is filled beyond the maximum mark,
excess anesthetic agent may spray from the filling
port when the bottle is removed.
023024
0
Vol.%
A
B
0
2
3
4
C
025
CAUTION
Health hazard
A bottle which is dropped may release significant
quantities of anesthetic agent, if broken.
Do not use a broken bottle and remove it from the
room.
0
2