Instructions for use D-Vapor/D-Vapor 3000 75
Alarms
Secondary audible alarm system
The secondary audible alarm system is only in
active operation when there is a mains power
failure (battery operation).
Suppressing the alarm tone
By using the Audio paused 2 min key (A),
medium priority alarms can be audibly suppressed
for 2 minutes. The yellow LED in the button lights
during this time. The alarm continues to be
indicated by the corresponding LED.
Alarm suppression can be deactivated by one of
the following actions:
The cause of the fault is resolved.
The control dial is set to the 0 position.
Lists of the causes and remedies for a fault, see
chapter "Problem solving", page 77.
Press the Audio paused 2 min key (A).
Alarm Cause Remedy
Continuous tone 7 s Mains supply failure; no further
battery operation possible.
Turn control dial to the 0 position. Change to a
vaporizer that does not require a mains supply.
Continuous tone Dosing request when the vaporiz-
er is not connected to the mains
power supply.
Turn control dial to the 0 position. Connect the
vaporizer to the mains power supply; wait for the
warm-up phase. Set the selected concentration.