Operation
72
Instructions for use D-Vapor/D-Vapor 3000
Storage
Storage for longer than 6 months
1 Draining the vaporizer, see page 69. Blowing
off the vaporizer, see page 71.
2 Press the 0 button and set the control dial to the
T position.
Plug-in adapters: engage locking lever in
control dial.
3 The vaporizer may be stored in any position.
4 Pack if necessary, see "Dispatch" on page 72.
Dispatch
1 Completely drain the vaporizer, see page 69.
Clean and disinfect the vaporizer, see page 89.
2 Engage control dial in the T position.
3 To dispatch anesthesia workstations, remove
the vaporizers if they are not permanently
connected.
4 Carefully pack each vaporizer separately! Use
the original packaging if possible. If the original
packaging is not available, use strong
packaging with at least 5 cm (2 in) of impact-
resistant material around each vaporizer.
Fasten the packaging securely.
WARNING
Internal damage to the vaporizer and incorrect
output concentrations
Internal damage to the vaporizer may occur if
the storage temperature range is exceeded.
Observe the storage temperatures, see
page 100.
Do not use the vaporizer otherwise. Have in-
spection carried out.
NOTE
Before putting into operation again, carry out in-
spection and test for operational readiness.
WARNING
Airfreight:
When transported as airfreight, liquid anes-
thetic agents and filled vaporizers are subject
to the IATA/ICAO Hazardous Goods Regula-
tions (UN 3334 "Aviation regulated liquid,
n.o.s. (anesthetic agent)", Class 9) if they have
been drained but not blown out.
NOTE
Road and sea transport:
When transported by road or sea, liquid anesthet-
ic agents and filled vaporizers are not subject to
any hazardous goods regulations unless national
regulations preclude it.