Operation
56
Instructions for use D-Vapor/D-Vapor 3000
Plug-in adapter S-2000 (D-Vapor only)
Only use white plug-in adapters for the D-Vapor.
1 Drop-in pin (A) on locking lever secure and
straight.
2 Both valve control pins (B) present.
3 Stop mechanism (C) undamaged, not bent.
4 Sealing surfaces (D) undamaged.
5 Interlock pins (E) undamaged, glide easily, and
cannot be removed.
6 Attachment plate (F) on the back of the
vaporizer present and secure.
ISO conical adapter (D-Vapor only)
Male cone connected to vaporizer inlet.
Female cone connected to vaporizer outlet.
Cones and sealing surfaces undamaged.
Permanent connection (D-Vapor only)
Vaporizer fixed securely to connector.
Filling systems
Baxter SAFE-FIL filling system for desflurane
(Suprane) and Piramal Fill filling system for
desflurane (Torrane)
Discoloration and contamination of the
anesthetic agent
If there is contamination shown by
discoloration, have the vaporizer repaired by
specialized service personnel.
Concentration check
Check the delivered concentration of anesthetic
agent weekly when continuous monitoring is not
available, see page 57.
Checking according to the checklist
Check the vaporizer according to the checklist,
see page 59.
Checking emergency power operation
Prerequisite: The battery has been charged for
1hour.
Set the flow to 4 L/min and the control dial to
6Vol%.
The vaporizer must dose for 5 minutes in battery
operation.