Problem solving
78
Instructions for use D-Vapor/D-Vapor 3000
Problems
Possible faults are listed in the following tables. The
tables will help to identify the underlying cause and
to quickly remedy it using the measures provided.
The different causes and remedies should be
worked through in the order listed until the fault has
been resolved.
Alarm cases
Priority Alarm indication Cause Remedy
Warning All LEDs flashing. Equipment fault Take equipment out of ser-
vice. Set control dial to the
0 position. Disconnect the
mains power supply. Re-
move the vaporizer from
the anesthesia workstation
if possible.
Repair by specialized ser-
vice personnel.
Warning Green Operational LED
flashes.
Red No Output LED flash-
es.
Dosing request during
warm-up phase
Malfunction
Set control dial to the 0 po-
sition. Wait until the green
Operational LED is lit con-
tinuously.
Warning Red No Output LED flash-
es.
Yellow Battery LED flash-
es.
End of battery operation Set control dial to the 0 po-
sition. Change to a vaporiz-
er that does not require a
mains supply.