Service
96
Instructions for use D-Vapor/D-Vapor 3000
5 Check of the filling level indication in the
T control dial position:
Tilt the vaporizer forward and backward. The
filling level indicated must fluctuate in the sight
glass.
If any of the test points are not satisfied, the vapor-
izer must not be operated and must be repaired by
specialized service personnel.
Maintenance
D-Vapor and D-Vapor 3000 do not need to undergo
preventive maintenance and do not need to be
regularly calibrated.
Repair
For all repairs, Dräger recommends DrägerService
and only the use of authentic Dräger parts.
Wearing parts
The following wear parts must be replaced if any
non-conformities with the specified values are
found during inspection and servicing or during
routine checks by the user.
Wear part Personnel responsible Note
Filter (gas inlet) Specialized service personnel
Mounting screw Specialized service personnel Replacement after unfastening
the connection adapter from the
vaporizer or unfastening the va-
porizer from its permanent con-
nection.
Locking pin (plug-in adapter) Specialized service personnel
O-ring (sealing plug) Service personnel Remove the sealing plug.
Remove faulty O-ring.
Fit new O-ring.
Battery Service personnel see page 38