EasyManua.ls Logo

Dräger D-Vapor

Dräger D-Vapor
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use D-Vapor/D-Vapor 3000 63
Operation
Adjusting concentration of anesthetic agent
First set the flow of fresh gas on the anesthesia
workstation.
Note regarding the illustration: To improve the representation,
the sight glass is shown here without any anesthetic agent.
If the control dial is set to the T position:
1 Press the 0 button (A) and turn the control dial
(B) counterclockwise to the required anesthetic
agent concentration. A short tone sounds.
For settings >12 Vol%:
Exceeding 12 Vol% is restricted by an interlock.
Press the 0 button (A) again and turn the
control dial (B) counterclockwise to an
anesthetic agent concentration >12 Vol%.
If the concentration is being reduced from
>12 Vol% to <12 Vol%, it is not necessary to press
the 0 button.
If the concentration cannot be set:
Do not force the control dial.
Check that all other vaporizers connected are in
the 0 or the OFF position and that Interlock it is
functional.
2 Check the filling level in the sight glass (C)
regularly. The filling level must be visible
between the minimum and maximum marks.
3 When the refill mark (D) has been reached, up
to 240 mL of desflurane may be refilled using an
anesthetic agent bottle.
4 Fill the vaporizer at the latest when the
minimum mark (E) is reached, see page 42.
NOTE
If fresh-gas flow is greater than 10 L/min, flow
noises may occur in the vaporizer. The noises are
not a result of leakage. Anesthetic gas is not being
released.
035
0
0
2
3
A
B
C
D
E
WARNING
Do not make any settings between the 0 and
the 2Vol% positions!
The concentration is not defined in this range.
WARNING
Risk of patient injury
If the anesthetic agent monitor shows implau-
sible values, check the vaporizer for incorrect
filling and the monitor for incorrect setting.

Table of Contents

Related product manuals