Instructions for use D-Vapor/D-Vapor 3000 13
For your safety and that of your patients
Additional safety information for
D-Vapor 3000
WARNING
Risk of patient injury
The transfer of data to the anesthesia worksta-
tion can be erroneous.
The indication of the control dial position on
the screen of an anesthesia workstation with
the Vapor View option is provided only for in-
formation.
Only rely on the concentration that is set on
the vaporizer.
WARNING
Risk of patient injury
The transfer of data to the anesthesia worksta-
tion can be erroneous.
The indication of the filling level on the screen
of an anesthesia workstation with the
Vapor View option is provided only for infor-
mation.
Only rely on the filling level shown on the va-
porizer.
WARNING
Risk of patient injury
The transfer of data to the anesthesia worksta-
tion can be erroneous.
The indication of the detected, active vaporiz-
er type on the screen of an anesthesia work-
station with the Vapor View option is provided
only for information.
Only rely on the anesthetic agent (color cod-
ing, labeling) shown on the vaporizer.