Instructions for use D-Vapor/D-Vapor 3000 11
For your safety and that of your patients
Product-specific safety information
WARNING
Risk of device errors
Various potentially dangerous situations can
arise that require the attentiveness of users.
Only use the device under the permanent su-
pervision of users in order to provide a
prompt remedy in the event of a malfunction.
WARNING
Not permitted modifications
Modifications to the medical device that are
not permitted may result in malfunctions.
This medical device must not be modified
without permission from Dräger.
WARNING
Risk of patient injury
Damage to the vaporizer can cause incorrect
output concentrations and serious harm to
the patient.
Handle the vaporizer with care. Be careful not
to tilt or drop. Do not carry by the control dial,
the control dial caps or the locking lever for
the plug-in adapter.
Do not use the vaporizer if it has been
dropped.
Check the vaporizer that has toppled over, see
"Checking operational readiness", page 54. If
the check is not possible, do not use the va-
porizer.
WARNING
Risk of patient injury
Incorrect concentrations of anesthetic agent
can harm the patient.
Use continuous monitoring of the delivered
concentrations of anesthetic agent, as per
ISO 21647 or ISO 80601-2-55 standard, with
upper and lower alarm limits to detect devia-
tions in concentration.
The operability of the measuring instrument
must be insured before you use it.
WARNING
Risk of patient injury
Incorrect output concentrations will not be de-
tected when operating from the external fresh-
gas outlet.
An anesthetic gas monitor must be used if the
vaporizer is used with anesthesia machine via
the external fresh-gas outlet.
WARNING
Risk of patient injury
Device function can be interfered with.
Do not use the vaporizer with magnetic reso-
nance imaging (MRI).
WARNING
Risk of patient injury
Deviation of the delivered concentration due
to filling the vaporizer with incorrect anesthet-
ic agent.
Only fill the vaporizer with desflurane.