Preparation
52
Instructions for use D-Vapor/D-Vapor 3000
Establishing electrical connections
Establishing the mains supply
Plug the mains plug into the power socket.
Establishing potential equalization
Establish potential equalization, e.g., for intra-
cardiac or intra-cranial surgery.
1 Remove the protective cap for the potential
equalization bonding stud (A).
2 Connect one end of the potential equalization
bonding conductor to the equipotential
equalization pin (B).
3 Connect the other end of the potential
equalization conductor to a central
equipotential equalization pin.
WARNING
Health hazard
Do not use mains plugs or power cables that
are damaged. There is a risk of electric shock
otherwise.
WARNING
Health hazard
Only connect the medical device to a mains
power supply with a protective ground. There
is a risk of electric shock otherwise.
NOTE
The mains plug must be readily accessible so that
the power supply to the device can be interrupted
in the event of a device failure.
033
A
B

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger D-Vapor and is the answer not in the manual?

Dräger D-Vapor Specifications

General IconGeneral
CompatibilityDräger anesthesia machines
Temperature CompensationAutomatic
Agent SpecificDesflurane
MountingSelectatec
Filling SystemEasy-Fill

Related product manuals