Instructions for use D-Vapor/D-Vapor 3000 51
Preparation
Setting the vaporizer which is not being used
1 Press the 0 button (A).
2 Engage control dial (B) at the T position.
Use of several vaporizers with ISO conical
adapters
Never connect the vaporizers in series!
031
0
0
0
0
A
B
B
A
032
WARNING
Danger to the patient from overdosing and/or
mixing anesthetic agents
Without the Interlock system there is a danger
that several vaporizers will be switched on
and operational at the same time.
If this happens, gas containing anesthetic
agent would flow from one vaporizer into the
next vaporizer, resulting in uncontrolled mix-
tures.
0 0