Operation
70
Instructions for use D-Vapor/D-Vapor 3000
5 Push the anesthetic agent bottle from below
into the filling opening. Anesthetic agent will
flow from the tank into the anesthetic agent
bottle.
6 Drain until the tank is empty or the bottle is full.
Stop pushing the anesthetic agent bottle from
below into the filling opening. Position the
vaporizer upright again. Press the locking
button and remove the anesthetic agent bottle.
To remove the 30 mL remaining in the
vaporizer, see "Blowing off the vaporizer"
on page 71.
7 If necessary, repeat the procedure with an
empty anesthetic agent bottle.
Ending the draining
1 Insert the sealing plug into the filling opening
and press it down until the lock engages
audibly.
2 Close the anesthetic agent bottle with the cap.
3 Mark the bottle:
"Used anesthetic agent."
Recommendation: Do not reuse.
Vaporizer with filling device for the
Piramal Fill filling system for desflurane
(Torrane)
Draining the vaporizer
Use an undamaged anesthetic agent bottle
provided by Piramal for desflurane (Torrane).
Unscrew the cap.
1 Remove the screw cap (A) from the filling port
by unscrewing.
2 Insert the anesthetic agent bottle (B) into the
filling port. Turn the bottle clockwise until a
spring resistance can be felt. Do not tilt the
bottle while turning it.
3 Turn the vaporizer upside down until the
anesthetic agent bottle is hanging vertically
downwards.
064
0
2
079080
0
Vol.%
A
0
Vol.%
B