112
Instructions for use D-Vapor/D-Vapor 3000
Correction of the flow measurement under the influence of desflurane
Hot wire volume monitors, which are not
compensated for desflurane, measure volumes
that are too great when the desflurane
concentration is high. See instructions for use for
the monitor used.
Correcting values for the minute volume MV
Correcting values for tidal volume VT
Example:
Desflurane concentration 8 Vol%
Displayed minute volume MV 10 L/min
MV
corrected = Displayed MV x
100 – 2 x desflu-
rane concentration
100
VT
corrected
= Displayed VT x
100 – 2 x desflu-
rane concentration
100
MV
corrected =10
x
100 – 2 x 8
100
MV
corrected = 10 x 0.84
=8.4L/min