Operation
64
Instructions for use D-Vapor/D-Vapor 3000
An anesthesia workstation may be moved with the
vaporizer switched on.
Changing the anesthetic agent
Set the vaporizer being used to the 0 position.
If only one vaporizer is connected or if
one of the connected vaporizers is to be
replaced
1 Disconnect the vaporizer, see page 67.
2 Connect the new vaporizer, see page 48.
3 Switch the anesthetic agent monitor to the new
anesthetic agent.
Two vaporizers with Interlock 2
1 Slide the latch of the Interlock 2 (A) into the
control dial on the vaporizer which has been
used until it engages.
2 On the vaporizer to be used, press the 0 button
and set the control dial (B) to the concentration
of anesthetic agent required.
WARNING
Concentration too low
If the underdosing alarm occurs:
Correct the setting, see "Influence of flow"
on page 110.
If the alarm continues:
Observe the anesthetic gas measurements
on the monitor. Set control dial to the 0 po-
sition. Disconnect the mains power supply.
Remove the vaporizer from the anesthesia
workstation. Contact specialized service
personnel.
WARNING
Concentration too low
During operation with both high fresh-gas
flow and high concentration, the actual deliv-
ered concentration may decrease, see "Influ-
ence of flow" on page 110.
WARNING
Jerky movements or tilting at an angle of more
than 10° can cause wrong dosage.
036
0
0
A
B