Operation
62
Instructions for use D-Vapor/D-Vapor 3000
Switching on and self-test
Plug the plug of the power cable into the power
socket. The self-test will start.
Self-test sequence
First a tone sounds briefly.
All 6 LEDs light for several seconds.
A tone then sounds briefly and 5 LEDs go out.
The green Operational LED (A) flashes,
indicating that the vaporizer is warming up.
After the warm-up phase (up to 5 min), a tone
sounds briefly. The green Operational LED (A)
lights continuously. If the vaporizer is cold (e.g.,
immediately after transport), the warm-up
phase will be extended.
The vaporizer is ready for use. The control dial can
be unlocked and the desired concentration set.
Alarm cases
If the control dial is operated while the green
Operational LED is flashing, an alarm tone sounds
and the red No Output LED (B) flashes. No
anesthetic agent will be dispensed.
If the green Operational LED is still flashing after
20 minutes' warm-up, contact specialized service
personnel, see "Alarm cases" on page 78.
If the yellow Fill Up LED (C) is lit, fill the vaporizer,
see "Filling the Vaporizer" on page 42.
WARNING
Health hazard
Only connect the device to a mains power
supply with a protective ground. There is a
risk of electric shock otherwise.
WARNING
Risk of patient injury
Do not use the vaporizer if one of the 6 LEDs
does not light during the self-test and the au-
dible signal does not sound. Repair by spe-
cialized service personnel.
Only use functioning vaporizers.
061
max.
min.
C
B
A