Instructions for use D-Vapor/D-Vapor 3000 45
Preparation
Ending the filling procedure
1 Stop pressing down the anesthetic agent bottle
(C).
Filling will be interrupted. The valve closes
automatically and prevents any loss of
anesthetic agent.
2 Wait 2 to 3 seconds while the anesthetic agent
runs into the tank.
3 Press the unlocking button and slowly withdraw
the anesthetic agent bottle from the filling
opening.
4 Check the filling level in the sight glass. The
vaporizer must be hanging or standing
vertically during the check.
The filling level must be visible and must not be
above the maximum mark.
5 Insert the sealing plug into the filling opening
and press it down until the lock engages
audibly.
6 Close the anesthetic agent bottle with the cap.
048
WARNING
Health hazard
Desflurane can spurt from the tank.
Wait 2 to 3 seconds while the anesthetic agent
runs into the tank.