Basic Menu (BSC)
The basic menu contains all of the configuration parameters necessary to make the transmitter operational.
Parameter Meaning Options Description
UNITS Measurement standard
ENGLSH (Inches)
METRIC (Millimeters)
The English/metric selection will also configure the transmitter to display sound speeds
in pipe materials and liquids as either feet per second (fps) or meters per second (mps),
respectively.
IMPORTANT: If the UNITS entry has been changed from ENGLSH to METRIC or from
METRIC to ENGLSH, the entry must be saved and the instrument reset (power cycled
or System Reset SYS RSET entered) in order for the transmitter to initiate the change in
operating units. Failure to save and reset the instrument will lead to improper transducer
spacing calculations and an instrument that may not measure properly.
ADDRESS
Modbus or BACnet
address
1…126
This address is for the EIA-485 port only. Ethernet addresses are set via the integrated
HTML application in the Ethernet Port. For transmitters ordered with a Modbus RTU
option, enter a value 1…126. For transmitters ordered with a BACnet MS/TP option, enter
a value 0…127. Each transmitter connected on the network must have a unique address
number assigned.
BAUD Baud rate of RS485
9600
14400
19200
38400
56000
57600
76800
BACNET ID BACnet device ID value 0…4194303 Applies to BACnet networks only.
XDCR MNT
Transducer mounting
method
V
W
Z
Selects the mounting orientation for the transducers based on pipe and liquid
characteristics. See Transducer Installation” on page15 .
XDCR HZ
Transducer transmission
frequency
500 kHZ
1 MHZ
2 MHZ
Transducer transmission frequencies are specific to the type of transducer and the size
of pipe. In general the DTTL 500 kHz transducers are used for pipes greater than 24
inches (600 mm). DTTR, DTTN and DTTH 1 MHz transducers, are for intermediate sized
pipes between 2 inches (50 mm) and 24 inches (600 mm). The DTTS and DTTC, 2 MHz
transducers, are for pipe sizes between 1/2 inch (13 mm) and 2 inches (50 mm)
FLO DIR
Transducer flow
direction
FORWARD
REVERSE
Allows the change of the direction the transmitter assumes is forward. When mounting
transmitters with integral transducers, this feature allows upstream and downstream
transducers to be electronically reversed making upside down mounting of the display
unnecessary.
PIPE OD Pipe outside diameter
ENGLSH (Inches)
METRIC (Millimeters)
Enter the pipe outside diameter in inches if ENGLSH was selected as UNITS; in millimeters if
METRIC was selected.
Charts listing popular pipe sizes have been included in the Appendix of this manual.
Correct entries for pipe O.D. and pipe wall thickness are critical to obtaining accurate flow
measurement readings.
PIPE WT Pipe wall thickness
ENGLSH (Inches)
METRIC (Millimeters)
Enter the pipe wall thickness in inches if ENGLSH was selected as UNITS; in millimeters if
METRIC was selected.
See “North American Pipe Schedules” on page110 for charts listing popular pipe sizes.
Correct entries for pipe O.D. and pipe wall thickness are critical to obtaining accurate flow
measurement readings.
PIPE MAT Pipe material
Select a material. This list is provided as an example. Additional pipe materials are added periodically. Select the
appropriate pipe material from the list or select OTHER if the material is not listed.
Acrylic ACRYLIC Glass Pyrex PYREX St Steel 304/316 SS 316
Aluminum ALUMINUM Nylon NYLON St Steel 410 SS 410
Brass (Naval) BRASS HD Polyethylene HDPE St Steel 430 SS 430
Carbon Steel CARB ST LD Polyethylene LDPE PFA PFA
Cast Iron CAST IRN Polypropylene POLYPRO Titanium TITANIUM
Copper COPPER PVC CPVC PVC/CPVC Asbestos ASBESTOS
Ductile Iron DCTL IRN PVDF PVDF Other OTHER
Fiberglass-Epoxy FBRGLASS St Steel 302/303 SS 303
Conguration
Page 36 May 2016TTM-UM-00136-EN-06

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the dynasonics TFX Ultra and is the answer not in the manual?

dynasonics TFX Ultra Specifications

General IconGeneral
Branddynasonics
ModelTFX Ultra
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals