VE-1010 Veterinary PC ECG User Manual EMC Information
- 68 -
Guidance and manufacture’s declaration - electromagnetic immunity - for all
EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration - electromagnetic immunity
VE-1010 VET PC ECG is intended for use in the electromagnetic environment specified below.
The customer or the user of VE-1010 VET PC ECG should assure that it is used in such an
environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
Not applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV line to line
2 kV line to ground
Not applicable Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
3A/m 3A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Not applicable Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
VE-1010 VET PC ECG
requires continued
operation during power
mains interruptions, it is
recommended that
VE-1010 VET PC ECG be
powered from an
uninterruptible power
supply or a battery.
NOTEU
T
is the a.c. mains voltage prior to application of the test level.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN INSTRUMENTS VE-1010 and is the answer not in the manual?

EDAN INSTRUMENTS VE-1010 Specifications

General IconGeneral
BrandEDAN INSTRUMENTS
ModelVE-1010
CategoryMedical Equipment
LanguageEnglish

Related product manuals