VE-1010 Veterinary PC ECG User Manual EMC Information
- 69 -
Guidance and manufacture’s declaration - electromagnetic immunity - for EQUIPMENT
and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration - electromagnetic immunity
VE-1010 VET PC ECG is intended for use in the electromagnetic environment specified below.
The customer or the user of VE-1010 VET PC ECG should assure that it is used in such an
environment.
Immunity
test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3V
rms
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of VE-1010
VET PC ECG, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Pd 2.1
Pd 2.1 80 MHz to 800 MHz
Pd 3.2 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN INSTRUMENTS VE-1010 and is the answer not in the manual?

EDAN INSTRUMENTS VE-1010 Specifications

General IconGeneral
BrandEDAN INSTRUMENTS
ModelVE-1010
CategoryMedical Equipment
LanguageEnglish

Related product manuals