F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance
11
WARNING
24 Parts and accessories used must meet the requirements of the applicable IEC 601
series safety standards, and/or the system configuration must meet the requirements
of the IEC 60601-1 medical electrical systems standard.
25 Connect the grounding wire to the equipotential grounding terminal in the main
system. If it is not evident from the instrument specifications whether a particular
instrument combination is hazardous or not, for example due to summation of
leakage currents, you should consult the manufacturer or an expert in the field, to
ensure that the necessary safety of all instruments concerned will not be impaired by
the proposed combination.
26 Electromagnetic Interference - Ensure that the environment in which the monitor is
installed is not subject to any source of strong electromagnetic interference, such as
CT, radio transmitters, mobile phone base stations, etc.
27 When installing the unit into a cabinet, allow for adequate ventilation, accessibility for
servicing, and room for adequate visualization and operation.
28 Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine,
dust area, high-temperature and humid environment.