F15 Series Fetal & Maternal Monitor Service Manual Periodic Tests
- 62 -
10.2.4 SpO2 Performance Test
This test checks the performance of SpO
2
measurement.
Required Equipment: SpO
2
Simulator
1) Connect the equipment and the monitor as shown below:
2) Switch on the monitor.
3) Generate the SpO2 signal 98% and 100% in turn.
4) Check if the SpO2 value displayed on the screen is accurate. Error within 2% is
acceptable.
10.2.5 NIBP Performance Test
This test checks the performance of NIBP measurement.
Required Equipment: Blood Pressure Simulator
1) Connect the equipment and the monitor as shown below:
2) Switch on the monitor.
3) Generate the NIBP signal 270/215 (235) and 40/10(20) in turn.
4) Check if the NIBP value displayed on the screen is accurate. Error within 8 mmHg is
acceptable.
10.2.6 TEMP Performance Test
This test checks the performance of TEMP measurement.
Required Equipment: Resistance Box
1) Connect the equipment and the monitor as shown below:
SpO
2
Sensor
SpO
2
Simulator
NIBP Cuff
Blood Pressure Simulator