EasyManuals Logo

EDAN F15 Series Service Manual

EDAN F15 Series
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
F15 Series Fetal & Maternal Monitor Service Manual Periodic Tests
- 62 -
10.2.4 SpO2 Performance Test
This test checks the performance of SpO
2
measurement.
Required Equipment: SpO
2
Simulator
1) Connect the equipment and the monitor as shown below:
2) Switch on the monitor.
3) Generate the SpO2 signal 98% and 100% in turn.
4) Check if the SpO2 value displayed on the screen is accurate. Error within 2% is
acceptable.
10.2.5 NIBP Performance Test
This test checks the performance of NIBP measurement.
Required Equipment: Blood Pressure Simulator
1) Connect the equipment and the monitor as shown below:
2) Switch on the monitor.
3) Generate the NIBP signal 270/215 (235) and 40/10(20) in turn.
4) Check if the NIBP value displayed on the screen is accurate. Error within 8 mmHg is
acceptable.
10.2.6 TEMP Performance Test
This test checks the performance of TEMP measurement.
Required Equipment: Resistance Box
1) Connect the equipment and the monitor as shown below:
SpO
2
Sensor
SpO
2
Simulator
NIBP Cuff
Blood Pressure Simulator

Table of Contents

Other manuals for EDAN F15 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F15 Series and is the answer not in the manual?

EDAN F15 Series Specifications

General IconGeneral
BrandEDAN
ModelF15 Series
CategoryMonitor
LanguageEnglish

Related product manuals