F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance
8
does not accept responsibility for safety compliance, reliability and performance if
modifications or repairs are carried out by unauthorized personnel. Identical replacement
parts must be used.
The typical operator's position is in front of the monitor. Please position the device in a
location where the operator can easily see the screen and access the operating controls.
The protective degree against electric shock of the patient connections is:
Ultrasound (FHR1, FHR2, FHR3)
External TOCO
Fetal Movement Mark (FM)
Fetal Stimulator (FS)
Intrauterine Pressure (IUP)
Maternal Electrocardiography (MECG)
Direct Electrocardiography(DECG)
Fetal Electrocardiography (FECG)
Type CF
Non-invasive Blood Pressure (NIBP)
Arterial Oxygen Saturation (SpO
2
)
Temperature (TEMP)
Type CF, defibrillation-proof
The monitor described in this service manual is not protected against:
a) The effects of high frequency currents
b) The interference of electrosurgery equipment
1.3 Safety Precautions
WARNING and CAUTION messages must be observed. To avoid the possibility of injury,
observe the following precautions during the maintaining the instrument.

Table of Contents