F15 Series Fetal & Maternal Monitor Service Manual Installation
- 5 -
Chapter 2 Installation
WARNING
The system installation should be operated by serviceman authorized by the manufacturer.
2.1 Environmental Requirements
Working Environment
Temperature:
0 ºC ~ + 40 ºC ( +32 ºF ~ +104 ºF)
Relative Humidity:
15% ~ 95% (non-condensing)
Atmospheric Pressure:
86 kPa ~ 106 kPa
Transport and Storage
Temperature:
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
Relative Humidity:
15% ~ 93% (non-condensing)
Atmospheric Pressure:
70 kPa ~ 106 kPa
Classification
Anti-electric Shock Type:
Class I equipment with internal power supply
Anti-electric Shock Degree:
MECG,FHR1,FHR2,FHR3
,TOCO,MARK,EXT.1,
DECG, IUP,FECG
CF
SpO2, NIBP , TEMP
CF (Defibrillating-proof)
Degree of Protection against
Harmful Ingress of Water:
Main Unit:
IPX2, protected against vertically falling water drops
when enclosure tilted up to 15o (provided recorder
drawer is shut and the monitor is not mounted on the
wall vertically)
US/TOCO/Fetal&Maternal Modules: IP68, dust-tight
and protected against the effects of continuous emersion
in water
FT20 Telemetry Transmitter: IPX2
Degree of Safety in Presence of
Flammable Gases:
Equipment not suitable for use in presence of
flammable gases
Disinfection/Sterilizing Method:
Refer to User Manual for details