EasyManuals Logo

EDAN F15 Series Service Manual

EDAN F15 Series
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
F15 Series Fetal & Maternal Monitor Service Manual Installation
- 5 -
Chapter 2 Installation
WARNING
The system installation should be operated by serviceman authorized by the manufacturer.
2.1 Environmental Requirements
Working Environment
Temperature:
0 ºC ~ + 40 ºC ( +32 ºF ~ +104 ºF)
Relative Humidity:
15% ~ 95% (non-condensing)
Atmospheric Pressure:
86 kPa ~ 106 kPa
Transport and Storage
Temperature:
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
Relative Humidity:
15% ~ 93% (non-condensing)
Atmospheric Pressure:
70 kPa ~ 106 kPa
Classification
Anti-electric Shock Type:
Class I equipment with internal power supply
Anti-electric Shock Degree:
MECG,FHR1,FHR2,FHR3
,TOCO,MARK,EXT.1,
DECG, IUP,FECG
CF
SpO2, NIBP , TEMP
CF (Defibrillating-proof)
Degree of Protection against
Harmful Ingress of Water:
Main Unit:
IPX2, protected against vertically falling water drops
when enclosure tilted up to 15o (provided recorder
drawer is shut and the monitor is not mounted on the
wall vertically)
US/TOCO/Fetal&Maternal Modules: IP68, dust-tight
and protected against the effects of continuous emersion
in water
FT20 Telemetry Transmitter: IPX2
Degree of Safety in Presence of
Flammable Gases:
Equipment not suitable for use in presence of
flammable gases
Disinfection/Sterilizing Method:
Refer to User Manual for details

Table of Contents

Other manuals for EDAN F15 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F15 Series and is the answer not in the manual?

EDAN F15 Series Specifications

General IconGeneral
BrandEDAN
ModelF15 Series
CategoryMonitor
LanguageEnglish

Related product manuals