F15 Series Fetal & Maternal Monitor Service Manual Connecting to OB TraceVue System
- 66 -
Appendix 2 Connecting to OB TraceVue System
Preparation
Prior to connecting F15/F15Air to the OB TraceVue obstetrical documentation and surveillance
system of Philips, make sure that:
1) Net Version of F15/F15Air on the Network Setup interface has this option: PHILIPS.
2) The OB TraceVue system of Philips has a network interface (DB9 or RJ45) available.
3) You have the proper connecting cable.
If a DB9 interface is available, you need a Serial Port Cable (01.13.107702):
Figure A2-1 Serial Port Cable
If an RJ45 interface is available, you need a Serial Port to RJ45 Cable (01.13.107974 or
01.13.036124). The part numbers are shown on the labels attached to the cables.
Figure A2-2 Serial Port to RJ45 Cable
Besides, you need a DB9 head Y shape cable