F15 Series Fetal & Maternal Monitor Service Manual Functional Checks
- 11 -
3.4 Printing Check
Press the PRINT key to start printing. Check if the recorder starts printing.
Enable the Print Self-Test function. Restart the monitor and verify if the recorder prints a
baseline.
Enter DEMO mode and print some traces and check if all the traces are clear on the paper.
If any failure is detected, refer to section 7.5 Recorder Failures for details.
NOTE:
Please make sure the paper loading and setting are correct before the printing starts.
3.5 Alarms Check
Stimulate a signal that is higher than the upper limit or lower than the lower limit to activate a
physical alarm. Disconnect one of the plugs off the monitor to activate a technical alarm. Check if
the audible alarms and visible alarms are working properly.
If any failure is detected, refer to section 7.7 Alarm Failures for defective details.