F15 Series Fetal & Maternal Monitor Service Manual Modules’ Malfunction Verification
- 41 -
Back of the Main control board
4) Compare the measurement results with the reference results in the list. If any one of the
results exceeds the reference range, the main control board defection is confirmed.
Replacement of the main control board is recommended.