Important
Whenever you change a discrete output parameter, verify all other discrete output parameters
before returning the meter to service. In some situations, the transmitter automatically loads a set of
stored values, and these values may not be appropriate for your application.
10.3.1 Configure Discrete Output Source
ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Source
Overview
Discrete Output Source controls which device condition or process condition is reported
via the discrete output.
Procedure
Set Discrete Output Source to the desired option.
The default setting for Discrete Output Source is Fault.
Options for Discrete Output Source
Options for Discrete Output SourceTable 10-4:
Option
Label
State
Discrete output volt-
ageProLink III Field Communicator
Enhanced Event 1–5 Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
ON Site-specific
OFF 0 V
Calibration in Progress Calibration in Progress Calibration in Progress
ON Site-specific
OFF 0 V
Fault (default)
Fault Indicator
Fault ON Site-specific
OFF 0 V
Important
This table assumes that Discrete Output Polarity is set to Active High. If Discrete Output Polarity is set
to Active Low, reverse the voltage values.
Integrate the meter with the control system