k. Set Mass Flow Rate Unit to the unit to be used for Mass Flow (Calculated).
5. (Optional) To configure Mass Flow (External) and Volume Flow (Calculated):
a. Choose Configure > Manual Setup > Inputs/Outputs > External Device Polling.
b. Choose an unused polling slot.
c. Set Poll Control to Poll as Primary or Poll as Secondary.
d. Set External Device Tag to the HART tag of the external measurement device.
e. Set Polled Variable to Mass Flow from Coriolis Meter.
f. Choose Configure > Manual Setup > Measurements > Optional Setup > External
Inputs > Configure External Inputs > Mass.
g. Set Mass Flow Source to Enable.
h. Choose Configure > Manual Setup > Measurements > Mass.
i. Set Mass Flow Rate Unit to the unit used by the external device.
j. Choose Configure > Manual Setup > Measurements > Gas Standard Volume.
k. Set GSV Flow Unit to the unit to be used for Volume Flow (Calculated).
8.4 Set up concentration measurement using the
Field Communicator
This section guides you through the tasks required to set up, configure, and implement
concentration measurement.
1. Enable the concentration measurement application using the Field Communicator
(Section 8.4.1)
2. Configure a concentration measurement matrix using the Field Communicator
(Section 8.4.2)
3. Select the active concentration matrix using the Field Communicator (Section 8.4.3)
8.4.1
Enable the concentration measurement application
using the Field Communicator
The concentration measurement application must be enabled before you can perform any
setup. If the concentration measurement application was enabled at the factory, you do
not need to enable it now.
1. Choose Overview > Device Information > Applications > Enable/Disable Applications.
2. Enable the concentration measurement application.
Configure process measurement using the Field Communicator
82 Micro Motion
®
Gas Specific Gravity Meters (SGM)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion SGM Series and is the answer not in the manual?

Emerson Micro Motion SGM Series Specifications

General IconGeneral
BrandEmerson
ModelMicro Motion SGM Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals