13.4.3 Perform temperature calibration using the
Field Communicator
Temperature calibration establishes the relationship between the temperature of the
calibration fluids and the signal produced by the sensor.
Prerequisites
The temperature calibration is a two-part procedure: temperature offset calibration and
temperature slope calibration. The two parts must be performed without interruption, in
the order shown. Ensure that you are prepared to complete the process without
interruption. You will need a low-temperature calibration fluid and a high-temperature
calibration fluid. You will not see the effect of the calibration until both the temperature
offset calibration and the temperature slope calibration are complete.
Important
Consult customer support before performing a temperature calibration. Under normal
circumstances, the temperature circuit is stable and should not need an adjustment.