5.6 Change the label for the active calibration
ProLink III Device Tools > Calibration Data
Field Communicator Configure > Manual Setup > Calibration Factors
Overview
You can change the default calibration labels.
Procedure
1. Select and delete the value in the Calibration x Label field.
2. Enter your preferred calibration label and press Apply.
5.7 Select the active calibration
ProLink III Device Tools > Calibration Data > Active Calibration
Field Communicator Configure > Manual Setup > Measurements > Gas Calibration > Active Calibration
Overview
The SGM can store up to four calibrations. The active calibration specifies the calibration to
be used for measurement.
Important
Do not confuse the active calibration with the calibration being performed. For example, you may
perform Calibration 4, and then choose to use Calibration 3 for measurement.
Procedure
Set Active Calibration to the calibration you want to use for measurement.
Related information
Review data for all calibrations
Purging and calibration
Configuration and Use Manual 47