EasyManua.ls Logo

Emerson Micro Motion SGM Series - B.4 Display Codes for Process Variables

Emerson Micro Motion SGM Series
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. To save the displayed value to transmitter memory, activate Scroll and Select
simultaneously and hold until the display changes.
If the displayed value is the same as the value in transmitter memory, you will be
returned to the previous screen.
If the displayed value is not the same as the value in transmitter memory, SAVE/
YES? flashes on the display. Activate Select.
6. Switch back from exponential notation to decimal notation.
a. Activate Select until the E is flashing.
b. Activate Select until d is displayed.
c. Activate Select.
B.4 Display codes for process variables
Display codes for process variablesTable B-2:
Code Definition
Standard
DENS Line Density
TEMP Line Temperature
EXT T Line Temperature (External)
EXT P Line Pressure (External)
Z Compressibility
MAG V Volume Flow Rate (External)
MAG M Mass Flow Rate (Calculated)
COR M Mass Flow Rate (External)
COR V Volume Flow Rate (Calculated)
DRIVE% Drive Gain
TP B Sensor Time Period
UCALC User-Defined Calculation Output
Concentration measurement
CONC Concentration
NET M Net Mass
NET V Net Volume
Gas measurement
BASE/DENS or BDENS Base Density
SG Specific Gravity
RD Relative Density
Using the transmitter display
180 Micro Motion
®
Gas Specific Gravity Meters (SGM)

Table of Contents

Related product manuals