The transmitter automatically filters out alerts with Alert Severity set to Ignore.
2. To acknowledge a single alert, check the Ack checkbox for that alert. To
acknowledge all alerts at once, click Ack All.
Postrequisites
To clear A010, A011, A012, A013 alerts, you must correct the problem,
acknowledge the alert, then repeat the calibration.
To clear A001, A002, A029, A031 alerts, you must correct the problem,
acknowledge the alert, then power-cycle the transmitter.
For all other alerts:
- If the alert is inactive when it is acknowledged, it will be removed from the list.
- If the alert is active when it is acknowledged, it will be removed from the list
when the alert condition clears.
Related information
Alert data in transmitter memory
12.3.2 View alerts using the Field Communicator
You can view a list containing all alerts that are active, or inactive but unacknowledged.
To view active or unacknowledged alerts, choose Service Tools > Alerts.
All active alerts and unacknowledged alerts are listed.
Note
Only Fault and Informational alerts are listed. The transmitter automatically filters out alerts
with Status Alert Severity set to Ignore.
To refresh the list, choose Service Tools > Alerts > Refresh Alerts.
Related information
Alert data in transmitter memory
12.3.3
Alert data in transmitter memory
The transmitter maintains three sets of data for every alert that is posted.
For each alert occurrence, the following three sets of data are maintained in transmitter
memory:
Alert List
Alert Statistics
Recent Alerts
Transmitter operation
Configuration and Use Manual 125