Alert data in transmitter memoryTable 12-1:
Alert data structure
Transmitter action if condition occurs
Contents Clearing
Alert List As determined by the alert status bits, a list of:
All currently active alerts
All previously active alerts that have not
been acknowledged
Cleared and regenerated with every transmit-
ter power cycle
Alert Statistics One record for each alert (by alert number)
that has occurred since the last master reset.
Each record contains:
A count of the number of occurrences
Timestamps for the most recent posting
and clearing
Not cleared; maintained across transmitter
power cycles
Recent Alerts 50 most recent alert postings or alert clearings Not cleared; maintained across transmitter
power cycles