SGM default variablesTable 4-2:
Default variable Output option A Output options B and C
Primary Variable (PV), mA1 Sample Temperature • Specific Gravity for Calibration Set 1
• Molecular Weight for Calibration Set
1
• Relative Density
Secondary Variable (SV),
mA2
Time Period B Sample Temperature
Tertiary Variable (TV) • Specific Gravity for
Calibration Set 1
• Molecular Weight for
Calibration Set 1
• Relative Density
Sensor Time Period
Quaternary Variable (QV) Drive Gain Drive Gain
4.2 Enable access to the off-line menu of the
display
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Display Menus > Offline Menu
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must
enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from
another tool.
4.3
Disable HART security
If you plan to use HART protocol to configure the device, HART security must be disabled.
HART security is disabled by default, so you may not need to do this.
Prerequisites
• Strap wrench
• 3 mm hex key
Introduction to configuration and commissioning
Configuration and Use Manual 23