a. Enter the values to be used for the constants: A, B, X, Y.
b. Enter the values to be used for a, b, c, d, e, and f.
For these terms:
You can specify a constant value.
You can specify a process variable. If you choose this, the current value of the
process variable is used in the calculation.
Important
User-defined calculations are performed using the meter's internal measurement units.
Therefore:
If a constant represents a process variable, you must enter its value in the internal
measurement units.
If a constant will be used to modify a process variable, you must use the internal
measurement units to derive the constant.
3. If you selected User-Defined Calculation 2:
a. Enter the values to be used for the constants: A, B, C.
b. Enter the value to be used for t.
For this term:
You can specify a constant value.
You can specify a process variable. If you choose this, the current value of the
process variable is used in the calculation.
Important
User-defined calculations are performed using the meter's internal measurement units.
Therefore:
If a constant represents a process variable, you must enter its value in the internal
measurement units.
If a constant will be used to modify a process variable, you must use the internal
measurement units to derive the constant.
4. Enter a label to be used for the output of the user-defined calculation (the new
process variable).
5. (Optional) Set up a reporting method for the new process variable.
The new process variable can be configured as a display variable or a HART variable,
or assigned to an mA output. It can also be read using digital communications.
Important
The output of the user-defined calculation is based on internal measurement units for process
variables. You may need to convert this value to the configured units before using it in your
application or process.