Status alerts, causes, and recommendations (continued)Table 14-2:
Alert number Alert title Possible causes Recommended actions
A003 No Sensor Re-
sponse
The transmitter is not re-
ceiving one or more basic
electrical signals from the
sensor.
Check the drive gain and the pickoff voltage. See
Section 14.22 and Section 14.23.
A004 Temperature
Overrange
The RTD resistance is out of
range for the sensor.
Check your process conditions against the values
reported by the device.
Verify temperature characterization or calibration
parameters.
Contact Micro Motion.
A006 Characteriza-
tion Required
Calibration factors have not
been entered, or the sensor
type is incorrect, or the cali-
bration factors are incorrect
for the sensor type.
Verify all of the characterization or calibration pa-
rameters. See the sensor tag or the calibration
sheet for your meter.
Contact Micro Motion.
A008 Density Over-
range
Applies only to the active
calibration.
If calibrated for specific
gravity or relative densi-
ty, the overrange is
greater than 3.0
If calibrated for molecu-
lar weight, the over-
range is greater than
86.7 g/mol
If other alerts are present, resolve those alert con-
ditions first. If the current alert persists, continue
with the recommended actions.
Check for foreign material in the process gas or
fluid, coating, or other process problems.
Verify all of the characterization or calibration pa-
rameters. See the sensor tag or the calibration
sheet for your meter.
Check the drive gain and the pickoff voltage. See
Section 14.22 and Section 14.23.
Perform Known Density Verification.
Contact Micro Motion.
A009 Transmitter Ini-
tializing/Warm-
ing Up
Transmitter is in power-up
mode.
Allow the meter to complete its power-up se-
quence. The alert should clear automatically.
If other alerts are present, resolve those alert con-
ditions first. If the current alert persists, continue
with the recommended actions.
Verify that the transmitter is receiving sufficient
power.
- If it is not, correct the problem and cycle pow-
er to the meter.
- If it is, this suggests that the transmitter has an
internal power issue. Replace the transmitter.
A010 Calibration Fail-
ure
Many possible causes.
This alert will not clear until
you correct the problem,
acknowledge the alert, and
repeat the calibration.
Ensure that your calibration procedure meets the
documented requirements, cycle power to the
meter, then retry the procedure.
Troubleshooting
Configuration and Use Manual 153