Connection to mA output terminalsFigure C-3:
A. PC
B. RS-232 to Bell 202 converter
C. 250–600
Ω
resistance
D. External power supply
E. Transmitter with end-cap removed
Note
This figure shows a serial port connection. USB connections are also supported.
The signal converter must be connected across a resistance of 250–600 Ω. The mA
output requires an external power supply with a minimum of 250 Ω and 17.5 V. See
the following figure to help determine the appropriate combination of voltage and
resistance. Note that many PLCs have a built-in 250-Ω resistor. If the PLC is powering
the circuit, be sure to take this into consideration.
Using ProLink III with the transmitter
200 Micro Motion
®
Gas Specific Gravity Meters (SGM)