Tip
Micro Motion recommends using the default settings for Alert Severity unless you have a specific
requirement to change them.
Procedure
1. Select a status alert.
2. For the selected status alert, set Alert Severity as desired.
Option Description
Fault
Actions when fault is detected:
• The alert is posted to the Alert List.
• Outputs go to the configured fault action (after Fault Timeout has expired, if
applicable).
• Digital communications go to the configured fault action (after Fault Time-
out has expired, if applicable).
• The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
• Outputs return to normal behavior.
• Digital communications return to normal behavior.
• The status LED returns to green.
Informa-
tional
Actions when fault is detected:
• The alert is posted to the Alert List.
• The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
• The status LED returns to green.
Ignore
No action
Status alerts and options for Status Alert Severity
Status alerts and Status Alert SeverityTable 9-1:
Alert number Alert title Default severity
User can reset
severity
A001 EEPROM Error
Fault
No
A002 RAM Error
Fault
No
A003 No Sensor Response
Fault
Yes
A004 Temperature Overrange
Fault
No
A006 Characterization Required
Fault
Yes
A008 Density Overrange
Fault
Yes
A009 Transmitter Initializing/Warming Up
Ignore
Yes
A010 Calibration Failure
Fault
No
Configure device options and preferences
Configuration and Use Manual 91