Status alerts and Status Alert Severity (continued)Table 9-1:
Alert number Alert title Default severity
User can reset
severity
A014 Transmitter Failure
Fault
No
A016 Sensor Temperature (RTD) Failure
Fault
Yes
A020 Calibration Factors Missing
Fault
Yes
A021 Transmitter/Sensor/Software Mis-
match
Fault
No
A029 Internal Electronics Failure
Fault
No
A030 Incorrect Board Type
Fault
No
A033 Insufficient Pickoff Signal
Fault
Yes
A037 Sensor Check Failed
Fault
Yes
A038 Time Period Signal Out of Range
Fault
No
A100 mA Output 1 Saturated
Informational
To Informational or
Ignore only
A101 mA Output 1 Fixed
Informational
To Informational or
Ignore only
A102 Drive Overrange
Informational
Yes
A104 Calibration in Progress
Informational
To Informational or
Ignore only
A106 Burst Mode Enabled
Informational
To Informational or
Ignore only
A107 Power Reset Occurred
Informational
Yes
A113 mA Output 2 Saturated
Informational
To Informational or
Ignore only
A114 mA Output 2 Fixed
Informational
To Informational or
Ignore only
A115 No External Input or Polled Data
Informational
To Informational or
Ignore only
A118 Discrete Output 1 Fixed
Informational
To Informational or
Ignore only
A120 Curve Fit Failure (Concentration)
Informational
No
A132 Sensor Simulation Active
Informational
Yes
A133 EEPROM Error (Display)
Informational
Yes
A136 Incorrect Display Type
Informational
Yes
Configure device options and preferences
92 Micro Motion
®
Gas Specific Gravity Meters (SGM)