6.3.1 Device status
The current device status is shown in Input Register 1000. You can view the device status
register by opening the Diagnostic window or the View Input Registers window.
Double-clicking the Value field of the Device Status register opens an expanded bitfield
window with information on the current device status as shown in Figure 6-5.
Figure 6-5: Input Register Device Status
A. Double-click the Value field to open the Expanded Bitfield window
Table 6-4: Device Status Messages
Message Bit no. Description Action
Device Warning 1 A device warning is active. See Warning messages for
details.
Device Error 7 A device error is active. See Error messages for details.
Simulation Mode
Active
8 Simulation mode active. Stop Simulation mode.
Write Protected 18 The device is write protected
with a switch or in the
TankMaster WinSetup
program.
Check the write protection
switch, see Write protection
switch.
Check write protection state in
TankMaster WinSetup, see
Write protection using
TankMaster
™
.
Related information
View and configure diagnostic registers using TankMaster
Viewing input and holding registers using TankMaster
Service and troubleshooting Reference Manual
December 2019 00809-0100-2410
106 Reference Manual