Liquiline System CA80SI Operation
Endress+Hauser 81
Menu/Setup/Outputs/Current output x:y
1)
Function Options Info
Current output Selection
Off
On
Factory setting
Off
Use this function to activate or deactivate a
variable being output at the current output
Source of data Selection
None
Connected inputs
Controller
Factory setting
None
The sources of data on offer depend on your
device version.
The main value of the analyzer and all the
sensors and controllers connected to inputs are
available for selection.
Measured value Selection
None
Depends on the Source
of data
Factory setting
None
The measured value you can select depends on
the option selected under Source of data .
The list of the dependent measured values can be found in the Measured value depending on the Source
of data→  81.
Apart from measured values from connected sensors, you can also select a controller as the data source.
The best way to do so is by using the Additional functionsmenu. Here, you can select and configure the
current output for outputting the controlled variable.
Range lower value Range of adjustment and
factory settings depending
on the Measured value
You can output the entire measuring range or
just some of it at the current output. To do so,
specify the upper and lower range values in
accordance with your requirements.
Range upper value
Hold behavior
(for sensors)
Selection
Freeze
Fixed value
None
Factory setting
Depends on the
channel:output
Freeze
The device freezes the last current value.
Fixed value
You define a fixed current value that is output at
the output.
None
A hold does not affect this current output.
Hold current
(for sensors)
Hold behavior = Fixed value
0.0 to 23.0 mA
Factory setting
22.0 mA
Specify which current should be output at
this current output in the hold state.
1) x:y = slot: output number
Measured value depending on the Source of data
Source of data Measured value
Selection
Main value
pH Glass Selection
Raw value mV
pH
Temperature
pH ISFET
ORP Selection
Temperature
ORP mV
ORP %